Call
Request a call

    Contact

    The DVC150 is ideal for characterizing display systems with a wide dynamic range and high measurement accuracy. Thanks to its compact dual monochromator, combined with a photomultiplier detector and integrated picoammeter, it achieves a remarkable dynamic range of six decades. This configuration ensures extremely accurate and reliable measurements. 

    The DVC150 is equipped with a direct-view telescope and fiber bundle, making it simple and convenient to operate. These features make it an ideal tool for professionals requiring precise, detailed analysis of high dynamic range display systems. 

    Characteristics : 

    Spectral range of operation

    380-780nm

    Spectral data interval

    1-5nm

    Spectral bandwidth (FWHM)

    1-5nm

    Wavelength accuracy

    ± 0.3nm, ± 0.05nm with software correction

    Wavelength reproducibility

    ± 0.05nm

    Resolution

    0.5nm

    Stray light rejection at 2.5 FWHM

    10-8

    Minimum working distance

    25mm with macro lens

    Maximum working distance

    50m with TL1 lens

    Bench space required

    1m deep x 2.5m wide typical including working space

    Services requirements

    2 x 110/220V AC mains sockets, 400VA total

    Télescopes :

    Lens

    Description

    Wavelength Range

    Focusing Range

    TEL301_LENS_TL-MACRO

    TL macro lens

    350-3000nm

    25mm

    TEL301_LENS_TL1-ACH-VIS

    TL1 VIS achromat lens

    Optimised for 350-800nm

    1m-infinity

    TEL301_LENS_TL2-ACH

    TL2 close-up achromat lens

    Optimised for 400-700nm

    150mm