The SpectroMaster 300 MAN is a spectro-goniometer using the 365-1014nm spectral range to accurately measure the refractive index of optical material.

The high precision of this goniometer spectrometer is achieved through special focusing optics, so that the telescope and collimator can be focused to achieve better collimation at the specific measurement wavelength.

The SpectroMaster® HR Compact Manual spectrometer uses a differential micrometric screw for maximum sensitivity and precision when manually adjusting the bezel angle. Thanks to sophisticated software support and alignment indicators, it has been proven that manual operators can achieve the same high accuracy as the motorized version, but at a lower measurement speed. All this allows a measurement accuracy of 5*10-6 on the refractive index of the sample.

 

SpectroMaster 300 MAN:

Maximum wavelength range 365-1014 nm
Standard Visible
Option NUV, NIR
Accuracy of angular measurement 0.5 arcsec
Measurement accuracy refractive index VIS : 5*10-6 (Contrôle température 0.1°)
Source HgCd
Detector CCD

 

Spectrometer collimator

  • Focusing Collimator
  • Clear aperture 50 mm
  • Focal length 300 mm

 

Autocollimator

  • Focusing Autocollimator
  • Clear aperture 50 mm
  • Focal length 300 mm
  • LED lighting
  • High resolution CCD camera

 

Spectral lamp housing used by the goniometer spectrometer

  • Single Spectral Lamp Housing (HgCd)
  • Spectral lamp power supply
  • Fused silica condenser optics

 

Sample holder

  • Sample table with 3-axis tilt adjustment screw (120°)
  • Linear circuit board

 

Spectral lamp housing used in spectro-goniometers

  • Single Spectral Lamp Housing (HgCd)
  • Spectral lamp power supply
  • Fused silica condenser optics

 

Sample holder

  • Sample table with 3-axis tilt adjustment screw (120°)
  • Precision manual linear plate for adjustment of the sample position
  • Precision manual rotary stage for sample rotation
  • Rotary encoder for measuring the sample rotation angle